Calibration acceptance criteria check

Paste the points from a calibration certificate, enter the instrument’s maximum permissible error (MPE) and see, point by point, whether it passes, how much margin is left and whether the lab’s uncertainty is small enough.

Certificate and criterion

One point per line: nominal value; instrument reading; U (expanded uncertainty). Paste straight from the certificate.
The limit your plant set for the instrument, without the ± sign.

Result

4/5points pass
Instrument fails
Most critical point
150 mm: 133% of the MPE
Lowest MPE/U ratio
3
Points in the zone of doubt
1

1 point(s) with the error inside the MPE but no room left for the uncertainty. The simple rule would pass them; the conservative rule does not. You cannot state that the instrument is within the limit.

The MPE/U ratio is 3 or more at every point. The usual ask is at least 3, and 4 is better.

NominalReadingError EUMPE|E| + UMPE/UMarginStatus
00.000.000.010.030.013+0.02Pass
2525.01+0.010.010.030.023+0.01Pass
5050.01+0.010.010.030.023+0.01Pass
100100.02+0.020.010.030.0330.00Pass
150150.03+0.030.010.030.043-0.01Doubt

How to check a calibration point by point

For each point on the certificate:

Error: E = instrument reading − nominal value

Conservative rule: pass if |E| + U ≤ MPE

Simple rule: pass if |E| ≤ MPE

MPE/U ratio (test uncertainty ratio, TUR): how many times the uncertainty fits in the MPE. Ask for 3 or more

The instrument passes only when every point passes. The MPE can be a fixed value, a percentage of reading (MPE × |nominal| ÷ 100) or a percentage of span (MPE × (end − start) ÷ 100).

The conservative rule uses the uncertainty as a guard band: the real acceptance limit becomes MPE − U. This is the idea behind ILAC-G8, the guideline on decision rules: when the result plus its uncertainty still fits the limit, you can state conformity. When the error fits the MPE but the error plus the uncertainty does not, the point is in the zone of doubt: you can say neither that it is in nor that it is out. ISO/IEC 17025 clause 7.8.6 requires the decision rule to be agreed beforehand whenever the lab states conformity.

Worked example

The values preloaded in the calculator are a 0 to 150 mm caliper, 0.01 mm resolution, with an MPE of ±0.03 mm and a calibration uncertainty of 0.01 mm at every point:

Nominal (mm) Reading Error E |E| + U Margin Status
0 0.00 0.00 0.01 +0.02 Pass
25 25.01 +0.01 0.02 +0.01 Pass
50 50.01 +0.01 0.02 +0.01 Pass
100 100.02 +0.02 0.03 0.00 Pass
150 150.03 +0.03 0.04 −0.01 Doubt
  1. At 100 mm, 0.02 + 0.01 = 0.03, exactly the MPE. It passes with no margin at all
  2. At 150 mm, the 0.03 mm error fits the MPE, but 0.03 + 0.01 = 0.04 does not
  3. The MPE/U ratio is 0.03 ÷ 0.01 = 3 at every point, right at the usual minimum
  4. Result: 4 of 5 points pass, the instrument fails under the conservative rule

Under the simple rule (|E| ≤ MPE) the same caliper would pass. This is exactly the case the conservative rule is for: given the lab’s uncertainty, nobody can guarantee the error at 150 mm is within ±0.03 mm.

How to read the result

  • Pass at every point: the instrument can go back into use until its next calibration.
  • Fail or doubt: the instrument should not measure parts until it is adjusted, repaired or downgraded to a use with a wider tolerance. And what it measured since its last good calibration needs a review: are the parts accepted near the limit still good?
  • Margin is what is left of the MPE at that point. Small margins at several points mean the instrument is close to the limit and may fail next time. Consider a shorter calibration interval.
  • An MPE/U ratio below 3 means the lab’s uncertainty is too large for the MPE you chose. Ask for a calibration with a smaller uncertainty or review the MPE. To see where an uncertainty comes from and how it is built, use the measurement uncertainty calculator.

Who sets the MPE

The MPE is the company’s decision, not the lab’s. It comes from the tolerance of what the instrument measures: a common rule is an MPE between 1/3 and 1/10 of the tolerance of the tightest characteristic. An MPE that is too loose lets through an instrument that will accept bad parts. One that is too tight fails instruments that would do the job. To see whether the process fits the tolerance, the Cp and Cpk calculator and the control chart complete the picture.

Common mistakes

  • Comparing only the error with the MPE and ignoring the uncertainty. A point can be out without you seeing it.
  • Taking the manufacturer’s specification as the MPE without thinking about the use. It is a starting point. What matters is the tolerance of the parts the instrument checks.
  • Ignoring the sign of the error. The rule uses |E|, but the sign tells you whether the instrument reads high or low, which helps with adjustment.
  • Passing on the average of the points. One failed point is enough to fail the instrument over the range it is used in.
  • Not reviewing what was measured with an instrument that came back failed. A failed calibration is where the analysis starts, not where it ends.

Frequently asked questions

What are the acceptance criteria for a calibration?

The most common in industry is the conservative rule: the instrument passes when, at every point, the absolute error plus the calibration uncertainty fits within the maximum permissible error, |E| + U ≤ MPE. A simpler rule compares the error alone, |E| ≤ MPE, but it gives no protection against the lab’s uncertainty.

What is maximum permissible error (MPE)?

The largest error you accept from an instrument for what it is used for. It can be a fixed value (±0.03 mm), a percentage of reading or a percentage of span. The company sets it from the tolerance of what the instrument measures, not the calibration lab.

How do you set the MPE for an instrument?

Start from the tightest tolerance the instrument will check. A common rule is an MPE between 1/3 and 1/10 of that tolerance: for a ±0.1 mm dimension (0.2 mm tolerance), a caliper with an MPE of ±0.02 to ±0.06 mm. Product standards and the manufacturer’s specification help settle the number.

Does the calibration lab pass or fail the instrument?

Usually not. The lab reports the error and uncertainty at each point. A statement of conformity only appears on the certificate if you ask for one and agree on the decision rule, as ISO/IEC 17025 requires (clause 7.8.6). Reviewing the certificate is the company’s job.

What is a good test uncertainty ratio?

The usual ask is a calibration uncertainty of at most 1/3 of the MPE (an MPE/U ratio of 3 or more), and 1/4 or better when you can get it. With a lower ratio, the uncertainty takes up so much of the MPE that good points end up in the zone of doubt.

What is a guard band in calibration?

A margin taken off the limit to account for measurement uncertainty. With the conservative rule the acceptance limit becomes MPE − U, so a point passes only if its error stays inside that narrower band. ILAC-G8 describes guard bands as one way to set a decision rule.